DC-5110/DC-5110RS
5040E51101110
13F., No.31, Lane 169, Kangning St., Xizhi Dist., New Taipei City 22180, Taiwan (R.O.C.)
886-2-2695-9688 886-2-2695-9693
Taichung Branch: 886-4-2319-7996 Kaohsiung Branch: 886-7-334-4103
* The specifications and appearance of the instrument are subject to change without notice.
Optional accessories:
1. D.O. sensor
2. Membrane kit
3. Electrolyte
4. Cleaning solution
5. Immersive protection holder 1M, 2M and 3M
6. Junction box
7. Extension cable
8. Auto clean kit
9. Stainless holder support base
Model DC-5110 DC-5110RS
Measuring mode % / ppm / mg/L / TEMP.
Measuring
range
% 0 ~ 600% (Depends on the sensor)
mg/L 0~60.00 mg/L (Depends on the sensor)
ppm 0~60.00 ppm (Depends on the sensor)
Temp. 0~140°C (Depends on the sensor)
Resolution
% 0.1 %
mg/L 0.01 mg/L / 0.001 mg/L
ppm 0.01 ppm / 0.001 ppm
Temp. 0.1℃
Accuracy
%
mg/L ±0.5%±1Digit
ppm
Temp. ±0.2°C±1Digit with temperature error correction function
Temperature Compensation NTC 30K or NTC 22K automatic compensation or manual adjustment
Salinity correction 0.0 ~ 45.0 ppt
Pressure correction 0.500 ~ 2.500 bar / 7.25 ~ 36.25 psi / 500~2500 mmHg manual adjustment
Ambient temperature 0~50°C
Storage temperature -10~70°C
Display
Large LCM with sensitization sensor for auto /
manual illumination function and contrast function
Text mode
Text mode
Chart mode
Trace mode
Analog output 1 Isolated DC 0/4~20mA corresponding to measurement, max. load 500Ω
Analog output 2
Isolated DC 0/4~20mA corresponding to
temperature, max. load 500Ω
-
Serial interface - RS-485 (Modbus RTU or ASCII)
Logbook - 50 event records
Setting
Contact 240VAC, 0.5A Max. (recommend)
Activate Hi/Lo, Hi/Hi, Lo/Lo selectable two limited programmable
Wash
Contact 240VAC, 0.5A Max. (recommend)
Time ON:0~99min. 59 sec. OFF:0~999 hours 59 min.
Voltage output DC±8V Max. 0.5W
Power supply 100~240VAC ±10%, 5W max., 50/60Hz
Installation Wall or Pipe or Panel mounting
Dimensions 96 mm × 96 mm × 132 mm (H×W×D)
Cut out dimensions 93 mm × 93 mm (H×W)
Weight 0.5Kg
Certification IP 65 (NEMA 4